Agilent Technologies 1989–2002
Agilent 3070 / 79000 Site Preparation
5-4
Chapter 5: Site Preparation – Environmental Requirements: Air Temperature Requirements
NOTE
The air temperature inside the system is not
considered stable until the system has been
powered for at least 30 minutes.
If the sensor finds that the air temperature inside the
system is too high, the system will shut down; typically,
the over-temperature shutdown point is 55 C (131 F).
See
on page 5-4 for information
on air conditioning requirements. See
page 5-3 for more temperature and humidity
specifications.
Cooling Requirements
Design the site cooling capability using the heat
dissipation estimates in
. The numbers are for
testheads with fully loaded modules, Serial Test power
supplies, and fully loaded support bays.
Table 5-2
System heat dissipation
Number of Modules
Dissipation
3X7X Series I Testhead (with
Support Bay)
1 Module
5100 BTU (1.5 kW)
2 Modules
8500 BTU (2.5 kW)
3 Modules
11900 BTU (3.5 kW)
4 Modules
15300 BTU (4.5 kW)
3X7X Series II, Series 3, including
3070PC, Testhead (with Support
Bay)
1 Module
10500 BTU (3.1 kW)
2 Modules
19700 BTU (5.8 kW)
3 Modules
28500 BTU (8.4 kW)
4 Modules
3700 BTU (10.9 kW)
327X /
79000
1 Module
10500 BTU (2.5 kW)
3X79CT POTS Bay
8 Channels
8500 BTU (2.5 kW)
44900A EFS Board Handler
1000 BTU (0.3 kW)