Agilent Technologies 1989–2002
Agilent 3070 / 79000 Site Preparation
6-8
Chapter 6: Site Preparation – Power Requirements: Power Requirements
Connecting Power to the PDU
Beginning with the serial numbers shown in
all systems are shipped with SO
1
power cords and plugs
factory-wired to the PDU.
One of two different power cords and plugs are attached
to the PDU depending on the country where the system
will be installed:
■
International
— E1135-61610: uses 5x6 mm
2
5-conductor cord with a IEC 60309 plug and
mating receptacle (see
■
North America
— E1135-61611: uses #10 AWG
5-conductor cord with NEMA plug that mates
with NEMA L21-30 (see
page 6-10). The female receptacle is available
locally.
1
Service with Oil resistant jacket
Table 6-2
Beginning serial numbers of systems shipped
with power cords
System Type
Testhead
Product
Testhead Serial
307X
E9900A
US38240533
317X
E9998A
US38240410
327X
E9997A
US38240248
79000
E2190A
US38050137