Chapter 9: Site Preparation – Receiving and Moving Instructions: Uncrating The System
If you have only small floor jacks, you may find it
easier to move the big pallet with one floor jack on
each end.
2
Place the pallet so that the system has room to roll off
to the side. The proper side of the system is where
you can see the "RAMP END" label. Leave three
meters (10 feet) of clearance at the rear of the pallet
for the system to roll off into. This clearance
provides sufficient room for the people who are
handling the system.
WARNING
✸
The straps are stretched tight and will whip when
cut. To avoid injury from flying straps, stand
clear of the straps when you cut them, and always
wear eye protection!
3
With diagonal cutters, cut the straps that hold the
cardboard box on the pallet.
4
Remove the plywood board from the top of the box.
Then remove the box.
NOTE
The Agilent Customer Engineer (CE) will
complete the remainder of this procedure.
5
Cut the interior strap on the ramp. After you have cut
the strap, the feet of the ramp should swing out. If
they do not, make sure the feet are perpendicular to
the ramp when you lower the ramp to the floor.
The bag labeled "See Inside" that contains
installation instructions also contains the hex wrench
or wrenches (5 millimeter for the testhead and 6
millimeter for the support bay if included) and the
9/16-inch open-end wrench.
NOTE
Save the installation procedure that is packed with
the system for the Agilent representative to use
when your system is installed.
6
With the 9/16-inch open-end wrench, remove the two
3.5-inch bolts that go through the bottom of the pallet
into the two-inch by six-inch board that holds the
ramp.
NOTE
Since the pallet feet only allow sufficient room to
remove the 3.5-inch bolts with a box or open-end
wrench, you should not attempt to remove these
bolts with a 9/16-inch socket wrench.
7
Make sure the ramp feet are in their proper positions,
and then lower the ramp to the ground.
8
Remove the plastic bags and foam padding from the
tester.
9
Retrieve the hex wrenches shipped with the system to
help you unbolt the shipping brackets from the tester.