Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure
Series 3 Current
Reduction
Conversion
Procedure
This section contains:
Current Reduction Conversion Preview
Determine the System Type and Power Option
Set the DUT Power Supply’s Input Voltage
Agilent 6621A and 6624A Line Voltage Settings
Agilent 6634A Line Voltage Settings
Agilent 6642A Line Voltage Settings
Rewire the Inputs to the Outlets
327X 1-Module System AC Outlets
79000 1-Module System AC Outlets
317X 2-Module System AC Outlets
307X 4-Module 1-PDU System AC Outlets
307X 4-Module 2-PDU System AC Outlets
E1085 POTS Bay and E2195 xDSL Bay AC Outlets
Connecting a Pigtail to the E1135C PDU
Current Reduction Conversion Preview
1
2
Determine the System Type and Power Option
3
4
Set the DUT Power Supply’s Input Voltage
.
5
Locate the system type and power option diagram
using
as shown for load balancing.
6
Reconnect the PDU to the Mains
. Note that if the
existing wiring is larger than 10 square millimeters
(#8 AWG), a pigtail should be used. See the section
Connecting a Pigtail to the E1135C PDU
NOTE
If a replacement E1135C is being installed and the
recommended wire size is greater than 10 square
millimeters (#8 AWG), which is the largest wire
size that will easily fit in the E1135C PDU, see the
section
Connecting a Pigtail to the E1135C
7
.
8
9
.