68
Window Trigger
The window trigger mode has a high trigger level and a low trigger level. The instrument triggers when
the input signal passes through either trigger level.
There are two kinds of window types: Absolute and Relative. They have different trigger level
adjustment methods. Under Absolute window type, the lower and the upper trigger levels can be
adjusted with the Level knob; under Relative window type, you instead set the center of the window,
and then a “delta” value. The upper trigger level is the center plus delta; the lower trigger level is
center minus delta.
l
If the lower and the upper trigger levels are both within the triggering range, the oscilloscope will
trigger on both rising and falling edges.
l
Suppose the upper trigger level is within the triggering range while the lower trigger level is out
of the triggering range. In that case, the oscilloscope will trigger on rising edge only.
To set absolute window triggers:
1.
Press
Trigger Setup
on the front panel to enter the TRIGGER function menu.
2.
Press the
Type
F key; then use the
Universal Knob
to select
Window
and push down the knob
to confirm.
3.
Press the
Source
F key; turn the
Universal Knob
to select CH1 or CH2 as the trigger source.
4.
Press the Window
Type
F key to select Absolute.
5.
Press the
Lower Upper
F key to select
Lower
or
Upper
trigger level; then press
Trigger Level
and turn the
Universal Knob
to adjust the position. The trigger level values are displayed at the
upper right corner of the screen.
The lower trigger level cannot be higher than the upper trigger level.
Содержание OSC-9200
Страница 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Страница 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Страница 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Страница 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Страница 101: ...86 l R W the type of a frame write or read l Data the value of the data Figure 42 I2C Decode List Display...
Страница 105: ...90...
Страница 116: ...101 l CRC the value of CRC Cyclic Redundancy Check field l ACK Acknowledgment bit Figure 57 CAN Decode List Displa...
Страница 135: ...120 Figure 75 Integral with Offset...
Страница 139: ...124 Figure 77 Measure Pulse Width...
Страница 146: ...131 Figure 83 Added the Measurement Note if the parameter does not match the measure condition it will display as...
Страница 190: ...175 DC Current Measurement AC Current Measurement...
Страница 204: ...info est aegis com www est aegis com...