![Aegis OSC-9200 Скачать руководство пользователя страница 181](http://html1.mh-extra.com/html/aegis/osc-9200/osc-9200_user-manual_2860410181.webp)
166
History
The history function can record the waveforms of the input channels until you press
Run/Stop
button.
In the Run state, the oscilloscope records input waveform repeatedly; when the memory fills up
(reaches the maximal frame), new frames will be recorded over the old frames. The latest frames will
be kept.
To use the History function, the HORIZONTAL
Format
must be set to
YT
.
To record and replay waveforms:
1.
Press
Shift
and
Meter
on the front panel to enable the History function.
l
If you are in the Run state, the waveform will enter the stop state.
l
If you are in the Stop state, and then enable the History function, the oscilloscope will keep
the stop state.
l
Press
Shift
and
Meter
again to turn off History.
2.
Press the
List
F key to turn on or off the list display. The list records the timestamp of every
frame. It is accurate to microseconds.
Figure 105 History
3.
Press the
Frame No.
F key; then turn the
Universal Knob
to select the frame to display.
l
The Frame format is A/B; A is the frames number that displays on the screen and B is the
maximum frame number you can set.
l
The maximum frame number is determined by the current sampling point (
Curr
value) and
sampling rate.
Содержание OSC-9200
Страница 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Страница 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Страница 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Страница 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Страница 101: ...86 l R W the type of a frame write or read l Data the value of the data Figure 42 I2C Decode List Display...
Страница 105: ...90...
Страница 116: ...101 l CRC the value of CRC Cyclic Redundancy Check field l ACK Acknowledgment bit Figure 57 CAN Decode List Displa...
Страница 135: ...120 Figure 75 Integral with Offset...
Страница 139: ...124 Figure 77 Measure Pulse Width...
Страница 146: ...131 Figure 83 Added the Measurement Note if the parameter does not match the measure condition it will display as...
Страница 190: ...175 DC Current Measurement AC Current Measurement...
Страница 204: ...info est aegis com www est aegis com...