152
Do Self Cal
The self-calibration program can quickly adjust the internal settings to give the most precise
measurement values. You can perform self-calibration at any time, but it’s especially important
when the temperature changes by 11° F (5° C) or more.
First, power on the scope and let it run for 30 minutes or more before the self-calibration. Then:
1.
Disconnect any probes from the scope.
2.
Press
Utility
on the front panel
,
and then press the
Do Self Cal
F key, and the oscilloscope will
pop-out the message box shown as below:
Figure 94 Do Self Cal
3.
Press the
F5
button on the front panel to perform the self-calibration program. During the
calibration, most of the keys are disabled.
4.
When the self-calibration program is finished, it will give the message, “
Press Run/Stop key to
exit
”. Press
Run/Stop
to exit the calibration interface.
Содержание OSC-9200
Страница 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Страница 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Страница 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Страница 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Страница 101: ...86 l R W the type of a frame write or read l Data the value of the data Figure 42 I2C Decode List Display...
Страница 105: ...90...
Страница 116: ...101 l CRC the value of CRC Cyclic Redundancy Check field l ACK Acknowledgment bit Figure 57 CAN Decode List Displa...
Страница 135: ...120 Figure 75 Integral with Offset...
Страница 139: ...124 Figure 77 Measure Pulse Width...
Страница 146: ...131 Figure 83 Added the Measurement Note if the parameter does not match the measure condition it will display as...
Страница 190: ...175 DC Current Measurement AC Current Measurement...
Страница 204: ...info est aegis com www est aegis com...