![Aegis OSC-9200 Скачать руководство пользователя страница 109](http://html1.mh-extra.com/html/aegis/osc-9200/osc-9200_user-manual_2860410109.webp)
94
UART Trigger and Serial Decode
Setup for UART Signals
1.
Press the
Shift
and
buttons to enter the
DECODE
function menu.
2.
Press the
Decode
F key and select the desired slot (Decode1 or Decode2).
3.
Press the
Protocol
F key and then select
UART
by turning
Universal Knob
.
4.
Press the
Signal
F key to enter the
SIGNAL
menu as below shows.
Figure 50 UART SIGNAL Menu
5.
Set RX:
a.
Press the
RX
to select the channel that is connected to the RX signal.
b.
Press the first
Threshold
key
to set the RX signal’s threshold voltage level by
Universal Knob
.
The threshold voltage level is for decoding, and it will be treated as the trigger voltage level
when you set the trigger type to serial.
6.
Set TX:
a.
Press the
TX
to select the channel that is connected to the TX signal.
c.
Press the second
Threshold
key
to set the TX signal’s threshold voltage level by
Universal
Knob
. The threshold voltage level is for decoding, and it will be treated as the trigger voltage
level when you set the trigger type to serial.
7.
Press the
F key to return to the previous menu.
8.
Press the
Configure
F key to enter the
BUS CONFIG
menu.
Figure 51 BUS CONFIG Menu
9.
Press the
Baud
F key to set baud rate.
l
The baud rate can be set as predefined value.
l
If the desired baud rate is not listed, press the
Baud
and select
custom
option, press the
Custom
and turn the
Universal Knob
to set the desired baud rate.
10.
Press the
Data Length
F key and set byte bits (5-8) by
Universal Knob
.
11.
Press the
Parity Check
F key to set the type of parity check (
Even, Odd, Mark, Space
or
None
).
12.
Press the
Stop Bit
F key to set the length of stop bit (
1, 1.5
or
2 bits
).
13.
Press the
Next Page
F key.
14.
Press the
Bit Order
F key to select the bit order (
LSB
or
MSB
).
15.
Press the
Idle Level
F key to set the idle level (
LOW
or
HIGH
).
Содержание OSC-9200
Страница 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Страница 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Страница 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Страница 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Страница 101: ...86 l R W the type of a frame write or read l Data the value of the data Figure 42 I2C Decode List Display...
Страница 105: ...90...
Страница 116: ...101 l CRC the value of CRC Cyclic Redundancy Check field l ACK Acknowledgment bit Figure 57 CAN Decode List Displa...
Страница 135: ...120 Figure 75 Integral with Offset...
Страница 139: ...124 Figure 77 Measure Pulse Width...
Страница 146: ...131 Figure 83 Added the Measurement Note if the parameter does not match the measure condition it will display as...
Страница 190: ...175 DC Current Measurement AC Current Measurement...
Страница 204: ...info est aegis com www est aegis com...