![Aegis OSC-9200 Скачать руководство пользователя страница 63](http://html1.mh-extra.com/html/aegis/osc-9200/osc-9200_user-manual_2860410063.webp)
48
Sequence Mode
Sequence is an acquisition mode that does not display the waveform during the sampling process.
This improves the waveform capture rate, and the maximal capture rate is 400,000 wfs/s. Because of
the high capture rate, sequence mode can capture small probability events effectively.
The oscilloscope runs and fills a memory segment for each trigger event. It repeats this process until
all memory segments are filled, and then displays the waveforms on the screen.
To use sequence mode, the HORIZONTAL Format must be set to
YT
.
Do the following steps to use the sequence mode.
1.
Press
Acquire
on the front panel to enter the ACQUIRE function menu;
2.
Press the
Sequence
F key to enter the SEQUENCE function menu.
Figure 23 SEQUENCE Function Menu
3.
Press the
Max. Segments
F key; and then turn the
Universal Knob
to select the desired value.
Do the following steps to replay the sequence waveform under history mode:
1.
Press the
Shift
and
Meter
button to enable HISTORY function.
Figure 24 HISTORY Function Menu
2.
Press the
List
F key to turn on the list display. The list records the acquisition time of every frame
and shows the frame number that displaying on the screen.
Содержание OSC-9200
Страница 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Страница 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Страница 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Страница 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Страница 101: ...86 l R W the type of a frame write or read l Data the value of the data Figure 42 I2C Decode List Display...
Страница 105: ...90...
Страница 116: ...101 l CRC the value of CRC Cyclic Redundancy Check field l ACK Acknowledgment bit Figure 57 CAN Decode List Displa...
Страница 135: ...120 Figure 75 Integral with Offset...
Страница 139: ...124 Figure 77 Measure Pulse Width...
Страница 146: ...131 Figure 83 Added the Measurement Note if the parameter does not match the measure condition it will display as...
Страница 190: ...175 DC Current Measurement AC Current Measurement...
Страница 204: ...info est aegis com www est aegis com...