![Aegis OSC-9200 Скачать руководство пользователя страница 7](http://html1.mh-extra.com/html/aegis/osc-9200/osc-9200_user-manual_2860410007.webp)
VI
Measurement Categories
The oscilloscope can measure in each category below within the listed voltage ranges. For the combination of
a probe assembly and an accessory, use the lower of the measurement categories’ limits.
AEGIS-OSC-9200
Max Input Voltage
Scope
(
BNC input
)
CAT II 300Vrms Between BNC Signal & Protective
Earth
CAT II 30Vrms Between BNC GND & Protective
Earth
CAT II 300Vrms Between BNC Signal & BNC GND
Scope
(
Probe input
)
CAT II 300Vrms Between Signal & Protective
Earth
CAT II 30Vrms Between Probe GND & Protective
Earth
CAT II 300Vrms Between Probe Signal & Probe
GND
Multimeter
CAT III 300Vrms, CATII 600Vrms.
SCD10A
[1]
CAT III 60Vrms.
SCD600MA
[1]
CAT III 60Vrms.
Note:
SCD10A is a 10A current adapter for the meter using to measure a current up to 10A, SCD600MA is a 600mA
current adapter for the meter using to measure a current up to 600mA. Refer to the DCI/ACI chapter for more
information.
Note:
Although measurements within a permanently installed motor would be Cat III, shaft voltage measurement
is considered Cat II, as neither the motor shaft or frame is directly energized under normal (non-fault) operation.
Содержание OSC-9200
Страница 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Страница 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Страница 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Страница 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Страница 101: ...86 l R W the type of a frame write or read l Data the value of the data Figure 42 I2C Decode List Display...
Страница 105: ...90...
Страница 116: ...101 l CRC the value of CRC Cyclic Redundancy Check field l ACK Acknowledgment bit Figure 57 CAN Decode List Displa...
Страница 135: ...120 Figure 75 Integral with Offset...
Страница 139: ...124 Figure 77 Measure Pulse Width...
Страница 146: ...131 Figure 83 Added the Measurement Note if the parameter does not match the measure condition it will display as...
Страница 190: ...175 DC Current Measurement AC Current Measurement...
Страница 204: ...info est aegis com www est aegis com...