![Aegis OSC-9200 Скачать руководство пользователя страница 152](http://html1.mh-extra.com/html/aegis/osc-9200/osc-9200_user-manual_2860410152.webp)
137
Persistence
With persistence, the oscilloscope updates the display with new acquisitions, but does not
immediately erase the results of previous acquisitions. All previous acquisitions are displayed with
reduced intensity. New acquisitions are shown in their normal color with normal intensity.
Do the following steps to set and clear persistence:
1.
Press
Display/Persist
on the front panel to enter the DISPLAY function menu.
2.
Press the
Persist
F key; then turn the
Universal Knob
to choose the setting:
l
Off
—turn of persistence.
l
Variable persistence time
(1 second, 5 seconds, 10 seconds, 30 seconds ) — how long prior
waveforms persist. Older waveforms are erased after this delay.
l
Infinite
—select “Infinite” Results of previous acquisitions are never erased. Use infinite
persistence to measure noise and jitter, see the worst-case extremes of varying waveforms,
look for timing violations, or capture rare events.
Figure 89 Persist Set to Infinite
3.
When
Persist
is
On
, press the
Clear Persist
F key to erase the results of previous acquisitions from
the display. The scope will start to accumulate acquisitions again.
4.
To return to the normal display mode, turn off Persist, and the previous acquisitions will be cleared
at once.
Содержание OSC-9200
Страница 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Страница 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Страница 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Страница 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Страница 101: ...86 l R W the type of a frame write or read l Data the value of the data Figure 42 I2C Decode List Display...
Страница 105: ...90...
Страница 116: ...101 l CRC the value of CRC Cyclic Redundancy Check field l ACK Acknowledgment bit Figure 57 CAN Decode List Displa...
Страница 135: ...120 Figure 75 Integral with Offset...
Страница 139: ...124 Figure 77 Measure Pulse Width...
Страница 146: ...131 Figure 83 Added the Measurement Note if the parameter does not match the measure condition it will display as...
Страница 190: ...175 DC Current Measurement AC Current Measurement...
Страница 204: ...info est aegis com www est aegis com...