![Aegis OSC-9200 Скачать руководство пользователя страница 107](http://html1.mh-extra.com/html/aegis/osc-9200/osc-9200_user-manual_2860410107.webp)
92
SPI Serial Decode
After completing the setup of SPI signal and trigger, we will decode SPI signals as follows.
1.
Press
Shift
and
→
Decode
. Select one of the options from the
Decode1
and
Decode2.
2.
Press the Configure F key to set the bit stream format and bits (4-32 bits)
3.
Press the
Display
F key and select
On
to display the result of decoding.
4.
Press the
Configure
F key to set bit order and data length.
5.
Press
the
List
F key to enter the
LIST
function menu.
Press the
Display
F key and choose the same options as the first step.
Press the
Scroll
F key and turn the
Universal Knob
to view all frames.
Press the
Lines
F key and set the number of lines by
Universal Knob
. The range of the lines
is 1 to 7.
Users can export the result list of the package to a USB storage device in csv format. Saving
the data (*.csv) is similar to the section "External save and recall”.
6.
Press the
Format
F key to change the character encoding format of the decoding’s result.
7.
Press the
Copy Setting
F key to enter the
COPY
function menu, to synchronize the
corresponding bus configuration and trigger configuration.
8.
Press the
Tips Info
F key to turn on or off the decoding limit prompt. When the decoding frame
number reaches the maximum, it will give an alert: "Decoding to maximum frame number
limitation!"
Interpreting SPI Decode
Decode Result Frames:
l
The data values are displayed in frames and are shown in white. The scope supports 4~96 bit
data display.
l
MISO — the decoding result of “Master-In, Slave-Out” line.
l
MOSI —the decoding result of “Master-Out, Slave-In” line.
l
Indicates there is not enough space on the display to show the complete content of a frame,
and some content is hidden.
Figure 48 SPI Decode Bus Display
Содержание OSC-9200
Страница 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Страница 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Страница 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Страница 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Страница 101: ...86 l R W the type of a frame write or read l Data the value of the data Figure 42 I2C Decode List Display...
Страница 105: ...90...
Страница 116: ...101 l CRC the value of CRC Cyclic Redundancy Check field l ACK Acknowledgment bit Figure 57 CAN Decode List Displa...
Страница 135: ...120 Figure 75 Integral with Offset...
Страница 139: ...124 Figure 77 Measure Pulse Width...
Страница 146: ...131 Figure 83 Added the Measurement Note if the parameter does not match the measure condition it will display as...
Страница 190: ...175 DC Current Measurement AC Current Measurement...
Страница 204: ...info est aegis com www est aegis com...