![Aegis OSC-9200 Скачать руководство пользователя страница 100](http://html1.mh-extra.com/html/aegis/osc-9200/osc-9200_user-manual_2860410100.webp)
85
I2C Serial Decode
After completing the setup of I2C signal and trigger, we will decode I2C signals as follows:
1.
Press
Shift
and
→
Decode
. Select one of the options from the
Decode1
and
Decode2.
2.
Press
Display
and select
On
to display the result of decoding.
3.
Press
Configure
to turn on or off the read or write bit.
4.
Press
List
to enter the
LIST
function menu.
Press
Display
and choose the same options as the first step.
Press
Scroll
and turn the
Universal Knob
to view all frames.
Press
Lines
and set the number of lines by
Universal Knob
. The range of the lines is 1 to 7.
When a packet of data is long, not all decoded data are displayed. Turn on the
Long Data
to
display the complete data.
Users can export the result list of the package to a USB storage device in csv format.
Saving the data (*.csv) is similar to the operation of setup files; see the section "External save
and recall" for details.
5.
Press
Format
to change the character encoding format of the decoding’s result.
6.
Press
Copy Setting
to enter the
COPY
function menu to synchronize the corresponding bus and
trigger configurations.
7.
Press
Tips Info
to turn on or off the decoding limit prompt. When the decoding frame number
reaches the maximum, it will give an alert: "Decoding to maximum frame number limitation!"
Interpreting I2C Decode
Decode Result Frames:
l
The address values are displayed at the beginning of a frame. The write address is displayed in
green, and the read address in yellow.
l
W/R bit is represented by (W) and (R), following the address value.
l
The data value is displayed in white.
l
"~A" after a data or address bits indicates no acknowledgement. For example, DB~A.
l
Indicates there is not enough space on the display to show the complete content of a frame,
and some content is hidden.
Figure 41 I2C Decode Bus Display
Decode Result Lists:
l
Time — the horizontal displacement between current frame and trigger position.
l
Address — the address of a frame.
Содержание OSC-9200
Страница 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Страница 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Страница 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Страница 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Страница 101: ...86 l R W the type of a frame write or read l Data the value of the data Figure 42 I2C Decode List Display...
Страница 105: ...90...
Страница 116: ...101 l CRC the value of CRC Cyclic Redundancy Check field l ACK Acknowledgment bit Figure 57 CAN Decode List Displa...
Страница 135: ...120 Figure 75 Integral with Offset...
Страница 139: ...124 Figure 77 Measure Pulse Width...
Страница 146: ...131 Figure 83 Added the Measurement Note if the parameter does not match the measure condition it will display as...
Страница 190: ...175 DC Current Measurement AC Current Measurement...
Страница 204: ...info est aegis com www est aegis com...