![Aegis OSC-9200 Скачать руководство пользователя страница 158](http://html1.mh-extra.com/html/aegis/osc-9200/osc-9200_user-manual_2860410158.webp)
143
9. To Default Key
Caution:
This function overwrites the settings that pressing
Default
restores.
After saving the settings with
To Default Key
, pressing
Default
will then restore the modified settings
instead of the original factory default settings. Do not use this function unless you know exactly what
the current settings are and you are certain that you want them as default. You will still be able to
recall the factory settings through Save/Recall
à
Recall
à
Factory Default (see the final section, Factory
Setup), but you will not be able to do this with
Default
.
With that in mind, to change the default settings:
After choosing the settings you want, press
Shift
and then
Cursors
. Push the
Save
F key, then the
Type
F key, choose
To Default Key
,
and then the
Press to Save
F key.
10. FileConverter
The mini tool converts stored binary files to CSV format for viewing with a spreadsheet program. It can
convert .bin files, sample logger .slg files, and measure logger .mlg files.
This is ideal when collecting large datasets. For a waveform frame with deep memory such as 12 Mpts,
to save directly as a CSV file will take long time and will occupy a large amount of memory on a USB
storage device. Instead we recommend saving as a .bin file and then converting it to CSV with
FileConverter.
Содержание OSC-9200
Страница 1: ...AEGIS Shaft Voltage Tester Handheld Digital Oscilloscope AEGIS OSC 9200...
Страница 18: ...3 Appearance and Dimensions mm Figure 1 Front View Figure 2 Side view...
Страница 44: ...29 Horizontal Control Contents u Horizontal Scale u Horizontal Delay u Roll Mode u Zoom Mode...
Страница 56: ...41 Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Страница 60: ...45 Figure 21 With Random Noise Normal Mode Figure 22 With Random Noise Average Mode...
Страница 101: ...86 l R W the type of a frame write or read l Data the value of the data Figure 42 I2C Decode List Display...
Страница 105: ...90...
Страница 116: ...101 l CRC the value of CRC Cyclic Redundancy Check field l ACK Acknowledgment bit Figure 57 CAN Decode List Displa...
Страница 135: ...120 Figure 75 Integral with Offset...
Страница 139: ...124 Figure 77 Measure Pulse Width...
Страница 146: ...131 Figure 83 Added the Measurement Note if the parameter does not match the measure condition it will display as...
Страница 190: ...175 DC Current Measurement AC Current Measurement...
Страница 204: ...info est aegis com www est aegis com...