Reinstall the operating head cover. Do one last freeze test with the oper-
ating head cover reinstalled; the freeze pulse characteristics are slightly
different with it in place.
If the freeze vibration is still not enough, do not proceed further without
obtaining Hot-Line
®
service as recommended in Chapter 5,
“
Obtaining
Service
”
.
If you succeed in obtaining enough freeze vibration, the stir vibration
amplitude must be readjusted as follows:
1. Press the
SETUP
soft key. If the Supervisor/operator keyswitch is in
the
Supervisor
position, the display will change to
“
Select Setup
Item
”
.
2. At
“
Select Setup Item
”
, press the
START
soft key. The display will
change to
“
Setup Menu #: 0
”
.
3. At
“
Setup Menu #: 0
”
, press the
4
soft key, then press
ENTER
.
The display will change to
“
Set Stir Amplitude
”
.
4. At
“
Set Stir Amplitude
”
, press the
START
soft key. The display
will change to
“
Amplitude = xxx
”
, with
“
xxx
”
being a value of
approximately 140.
Looking from the side of the instrument, you should be able to see a stir
vibration blur at the tip of the stir/freeze wire. For 2-mL to 2.5-mL
cryoscope operation, the stir vibration blur at the very tip of the stir/
freeze wire should be between 1/8
”
and 1/4
”
or 3 and 6 mm. A small
amount of stirring is necessary to minimize temperature gradients in
viscous samples but more than 1/4
”
or 6 mm can cause poor repeatabili-
ty or frothing which may make it impossible to reliably determine the
freezing point.
If the stir vibration amplitude is not correct, press
>
to increase the stir
amplitude and the
“
xxx
”
value or press
<
to decrease the stir amplitude
and the
“
xxx
”
value. Continue to increase or decrease the
“
xxx
”
value as
necessary, until the stir vibration blur at the very tip of the stir/freeze
wire is between 1/8
”
and 1/4
”
or between 3 and 6 mm.
64
Advanced
®
Cryoscope Model 4250 User’s Guide
Содержание 4250
Страница 1: ...4255 Rev11 032317 The Advanced Cryoscope Model 4250 User s Guide...
Страница 6: ...Notes vi Advanced Cryoscope Model 4250 User s Guide...
Страница 14: ...xiv Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 28: ...Advanced Cryoscope Model 4250 User s Guide 6 Figure 3 Load Printer Paper...
Страница 56: ...34 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 60: ...Advanced Cryoscope Model 4250 User s Guide 38 Notes...
Страница 88: ...66 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 102: ...80 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 106: ...Advanced Cryoscope Model 4250 User s Guide 84 Notes...
Страница 109: ...87 Supplemental RS 232 Information Null Modem Cable Shield connects to metalized con nector housing at both ends...
Страница 110: ...88 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 114: ...92 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 118: ...Advanced Cryoscope Model 4250 User s Guide 96 Date Problem Symptom Action...