11. Replace the head cover and its two attaching screws.
12. Reconnect the power cord and turn the instrument on.
13. Make sure the
Supervisor/Operator keyswitch
is in the
Supervisor
position while using the
SETUP
and
CALIB
menus
(see Figure 7 and Chapter 2
“
Supervisor/Operator Keyswitch
”
).
14. When the operating head rises and
“
Press START to Continue
”
begins to scroll across the display, press the
START
soft key.
15. When the display changes to
“
Cryoscope Ready
”
, press the
TEST
soft key. The display will change to
“
Select Test Item
”
.
16. At
“
Select Test Item
”
, repeatedly press the
>
soft key until the dis-
play reads
“
Probe Bin Test
”
.
17. At
“
Probe Bin Test
”
, press
START
. The display will change to
“
[Start] Ready?
”
.
18. At
“
[Start] Ready?
”
, remove the empty sample tube from the freez-
ing chamber, pipette or measure a sample of probe bin setting fluid
into a clean sample tube, place the tube in the freezing chamber and
press the
START
soft key. The 4250 will run a special freezing
point test to determine the sample probe resistance and bin number.
NOTE
In emergency, H
2
O may be used in place of bin-setting fluid for
this test. The bin number thus determined will be acceptable for
bin-setting though the probe resistance displayed is that at the
freezing point of the H
2
O sample used, and will be different
from that for bin-setting fluid.
19. Record the sample probe resistance and bin number in the Service
Log at the end of the user's manual; then press the
STOP
soft key.
The display will change to
“
Probe Bin Test
”
.
20. Press
SETUP
to enter the
SETUP
menu, then repeatedly press the
>
soft key until the display reads
“
Set Sample Bin #
”
.
21. At
“
Set Sample Bin #
”
, press the
START
soft key. The display will
report the current sample bin number setting.
Advanced
®
Cryoscope Model 4250 User’s Guide
58
Содержание 4250
Страница 1: ...4255 Rev11 032317 The Advanced Cryoscope Model 4250 User s Guide...
Страница 6: ...Notes vi Advanced Cryoscope Model 4250 User s Guide...
Страница 14: ...xiv Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 28: ...Advanced Cryoscope Model 4250 User s Guide 6 Figure 3 Load Printer Paper...
Страница 56: ...34 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 60: ...Advanced Cryoscope Model 4250 User s Guide 38 Notes...
Страница 88: ...66 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 102: ...80 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 106: ...Advanced Cryoscope Model 4250 User s Guide 84 Notes...
Страница 109: ...87 Supplemental RS 232 Information Null Modem Cable Shield connects to metalized con nector housing at both ends...
Страница 110: ...88 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 114: ...92 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 118: ...Advanced Cryoscope Model 4250 User s Guide 96 Date Problem Symptom Action...