![Advanced Instruments 4250 Скачать руководство пользователя страница 52](http://html2.mh-extra.com/html/advanced-instruments/4250/4250_user-manual_3047840052.webp)
8. Dis/Enable Xon/Xoff
This menu item enables or disables the Xon/Xoff
“
handshaking
”
proto-
col for RS-232 communication (CTRL-”S” = STOP/ CTRL-”G” = GO).
1. At
“
Dis/Enable Xon/Xoff
”
, press
START
to display
“
[on] <
Xon/Xof * [off]
”
. The current setting is indicated by the symbol,
“
*
".
2. At
“
[on] < Xon/Xof * [off]
”
, press
<
or
>
to enable or disable ser-
ial-port Xon/Xoff handshaking.
3. Press
ENTER
to store the new setting (or
STOP
to restore the
original setting).
9. Serial Number
At
“
Serial Number
”
, press
START
to display the instrument serial
number (
“
Serial #: 123
”
). Press
STOP
to return to the setup menu. This
number is set by the factory and cannot be changed by the operator.
10. Select Plateau Mode
This menu item enables changing the way the 4250 evaluates the
plateau. An idealized, dimensionless freezing point plateau is illustrated
and described in Figure 1.
1. At “
Select Plateau Mode
”, press
START
to display the current
mode.
2. Press
<
or
>
to toggle between plateau-seeking and timed-plateau
detection modes.
3. Press
ENTER
to store the new setting (or
STOP
to restore the
original setting).
11. Set % DFB
This menu item enables storing a
“
base
”
freezing point in the instrument
memory and displaying, with each sample test result, the percent differ-
ence between the test result and the stored base.
30
Advanced
®
Cryoscope Model 4250 User’s Guide
Содержание 4250
Страница 1: ...4255 Rev11 032317 The Advanced Cryoscope Model 4250 User s Guide...
Страница 6: ...Notes vi Advanced Cryoscope Model 4250 User s Guide...
Страница 14: ...xiv Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 28: ...Advanced Cryoscope Model 4250 User s Guide 6 Figure 3 Load Printer Paper...
Страница 56: ...34 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 60: ...Advanced Cryoscope Model 4250 User s Guide 38 Notes...
Страница 88: ...66 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 102: ...80 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 106: ...Advanced Cryoscope Model 4250 User s Guide 84 Notes...
Страница 109: ...87 Supplemental RS 232 Information Null Modem Cable Shield connects to metalized con nector housing at both ends...
Страница 110: ...88 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 114: ...92 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 118: ...Advanced Cryoscope Model 4250 User s Guide 96 Date Problem Symptom Action...