bration test results are not acceptably repeated throughout the six
tests at each calibration level, the instrument will report
“
Calibration out of Range
”
and the entire calibration procedure
must be redone. If the repeatability problem recurs despite good
operating technique, please call or write immediately for Hot-Line
®
service.
NOTE
If the instrument has lost all previous calibration data and you
see the message
“
Recalibration Needed
”
, pay close attention to
the displayed prompts. You will be requested to run a slightly
different sequence than described above.
Advanced
®
Cryoscope Model 4250 User’s Guide
42
Содержание 4250
Страница 1: ...4255 Rev11 032317 The Advanced Cryoscope Model 4250 User s Guide...
Страница 6: ...Notes vi Advanced Cryoscope Model 4250 User s Guide...
Страница 14: ...xiv Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 28: ...Advanced Cryoscope Model 4250 User s Guide 6 Figure 3 Load Printer Paper...
Страница 56: ...34 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 60: ...Advanced Cryoscope Model 4250 User s Guide 38 Notes...
Страница 88: ...66 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 102: ...80 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 106: ...Advanced Cryoscope Model 4250 User s Guide 84 Notes...
Страница 109: ...87 Supplemental RS 232 Information Null Modem Cable Shield connects to metalized con nector housing at both ends...
Страница 110: ...88 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 114: ...92 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 118: ...Advanced Cryoscope Model 4250 User s Guide 96 Date Problem Symptom Action...