![Advanced Instruments 4250 Скачать руководство пользователя страница 53](http://html2.mh-extra.com/html/advanced-instruments/4250/4250_user-manual_3047840053.webp)
1. At
“
Set % DFB
”
, press
START
to display
“
Base Reading = xxx
”
.
The freezing point units will be the same as the units selected for
the regular freezing point test displays.
2. Press
CLEAR
to set the base to 0. Then use the numeric keypad
to enter the base value required.
3. When the base is set as required, press
ENTER
to save the setting
and return to the display to
“
Set % DFB
”
.
4. At
“
Set % DFB
”
, press
STOP
to return to the display
“
Cryoscope
Ready
”
.
12. Select Display mH/mC
This menu item enables changing the readout temperature units (see
Foreword for a discussion of the units of measurement used for milk
cryoscopy).
CAUTION
Changing this setting requires changing the % DFB set-
ting (if not set at zero), as well. (See
Set % DFB
, above.)
Changing this setting will erase all data stored in Recall and
Statistics memory.
1. At “
Select Display mH/mC
”, press
START
to display “
[mH] <
Display * [mC]
”.
2. At “
[mH] < Display * [mC
]”, the current setting is indicated by
the symbol “*”. Press < or > to toggle the display readout
between mH and mC.
3. Press
ENTER
to store the new setting (or
STOP
to restore the
original setting).
13. Select Language
This menu item enables selecting one of the alternate language display-
panel messages.
31
Instrument Operation
Содержание 4250
Страница 1: ...4255 Rev11 032317 The Advanced Cryoscope Model 4250 User s Guide...
Страница 6: ...Notes vi Advanced Cryoscope Model 4250 User s Guide...
Страница 14: ...xiv Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 28: ...Advanced Cryoscope Model 4250 User s Guide 6 Figure 3 Load Printer Paper...
Страница 56: ...34 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 60: ...Advanced Cryoscope Model 4250 User s Guide 38 Notes...
Страница 88: ...66 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 102: ...80 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 106: ...Advanced Cryoscope Model 4250 User s Guide 84 Notes...
Страница 109: ...87 Supplemental RS 232 Information Null Modem Cable Shield connects to metalized con nector housing at both ends...
Страница 110: ...88 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 114: ...92 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 118: ...Advanced Cryoscope Model 4250 User s Guide 96 Date Problem Symptom Action...