![Advanced Instruments 4250 Скачать руководство пользователя страница 69](http://html2.mh-extra.com/html/advanced-instruments/4250/4250_user-manual_3047840069.webp)
Monthly
. Maintenance you should perform:
Clean or replace the air filter (Figure 5, item 7) monthly; replace as
necessary. A dirty filter affects the ability of the fan to cool internal
components; a rise in internal temperature may adversely affect relia-
bility.
Shutdown and storage
Overnight or weekend:
1. Rinse the probe clean with distilled water and wipe it dry.
2. Turn off the instrument, if desired.
3. Do not cover the instrument unless it is turned off.
One week or longer:
1. Rinse the probe clean with distilled water and wipe it dry.
2. Turn off the instrument.
3. Unplug the power cord from the wall outlet.
4. Cover the instrument.
Troubleshooting checks
Check operational requirements.
If you are experiencing difficulties
with your instrument, first carefully review the operational requirements
listed in the product specifications and the recommended setup and oper-
ating procedures. Check the heat transfer fluid level. Make sure that the
heat transfer fluid pump is fully primed.
Check fuses.
You will find the power switch and fuse holder beside the
power cord connector on the back panel of the instrument. Switch the
power switch to the off position and disconnect the power cord. Use a
small flat-bladed screwdriver or similar tool to pry out the fuse holder.
Visually check for a blown fuse. If there is any doubt, test the fuses with
a continuity checker or ohmmeter or simply replace them with new fuses.
Check probe and stir wire.
Run the A/D Test and the Probe Bin Test to
evaluate the probe operation. Refer to the Mandrel, Probe, and
47
Troubleshooting & Service
Содержание 4250
Страница 1: ...4255 Rev11 032317 The Advanced Cryoscope Model 4250 User s Guide...
Страница 6: ...Notes vi Advanced Cryoscope Model 4250 User s Guide...
Страница 14: ...xiv Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 28: ...Advanced Cryoscope Model 4250 User s Guide 6 Figure 3 Load Printer Paper...
Страница 56: ...34 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 60: ...Advanced Cryoscope Model 4250 User s Guide 38 Notes...
Страница 88: ...66 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 102: ...80 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 106: ...Advanced Cryoscope Model 4250 User s Guide 84 Notes...
Страница 109: ...87 Supplemental RS 232 Information Null Modem Cable Shield connects to metalized con nector housing at both ends...
Страница 110: ...88 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 114: ...92 Advanced Cryoscope Model 4250 User s Guide Notes...
Страница 118: ...Advanced Cryoscope Model 4250 User s Guide 96 Date Problem Symptom Action...