4/4
B 70
J 14 remote control plug connection
G 01059 - Edition 07 - Januar
y 10
Contact Function
S7
(30-31)
Authorization to
close or not the
inlet valve with
the key “
AUX
“ of
hand-held remote
control
Contact closed, authorization to close the inlet valve with the key
“
AUX
“ of hand-held remote control.
Contact open, no authorization to close the inlet valve with the key
“
AUX
“ of hand-held remote control.
This function is only valid when “
INLET VALVE OPTION
“ is “
VALID
“ in
the DEFINITION menu of hand-held remote control. This function is not
related to contact S6 and enables to function independently in local
mode or remote control mode.
Control contacts
(inputs) (continued)
Available dry contacts 50V - 1A.
Available outputs
These contacts open in
the presence of a fault
(normally closed):
Contact Function
1-34
Pump fail (pump running, valve opened, no alert)
2-3
Alert presence
4-5
Alarm presence - Unit stoppage
6-7
Maintenance alert
8-9
Pump running, correct purge and inlet valve opened
14-15
Pump operation (ADP or ADS)
16-17
State of inlet valve
35-36
Motor temperature alert
37-38
Purge alert
39-40
Exhaust pressure alert
These contacts close in
the presence of a fault
(normally open):
Contact Function
10-11
Alarm presence - Unit stoppage
12-13
Alert presence (without maintenance)
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two