G 03218 - Edition 01 - December 06
1/3
D 400
Pump or Roots does not start
Mains presence light indicator does not switch on
..
page 1
Hand-held remote control display is not lighting up
.
page 1
Display is showing a message or an alarm
..........
page 2
Incident
Cause
Troubleshooting
Mains presence light indicator
does not switch on
No mains
Check power supply.
Main switch on position “0”
Switch it on “
1
”.
External emergency stop engaged
Check the presence of cover plug
J1
,
or check its locking.
Unit emergency stop engaged
Rotate to unlock the emergency stop
on the front panel display.
CB1 breaker switched off
Check the amperage setting
(see
D 10
).
Switch it on.
CB2 breaker switched off
Switch it on (see
D 10
).
Other problem
Call customer service center.
Incident
Cause
Troubleshooting
Hand-held remote control display
is not lighting up, but the “mains
presence” green light indicator is on
CB3 breaker switched off
Switch it on (see
D 10
).
Twisted cable incorrectly connected
Check the connection.
Other problem
Call customer service center.
Mains presence light indicator does not switch on
Hand-held remote control display is not lighting up
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two