4/7
D 500
The pump is running and then is showing
a warning or fault message
G 03219 - Edition 01 - December 06
Incident
Cause
Troubleshooting
Fault
A300 functionnal block does no run
at ultimate speed
Mechanical problem
y
Check the exhaust line.
Other problem
Call customer service center.
Warning
No water flow
y
Check the water supply.
y
Check the main valve connection on
“Water valve“
connector (see
D 10
).
Insufficent water flow
y
Check that the main valve, the
flowmeter and the pipes are not
clogged.
y
Check the setting value of the
flowmeter (see
B 30
).
y
Check that there are no leaks.
Sensor not wired
y
Check the flowmeter connection on
the
“Waterflow“
connector
(see
D 10
).
Other problem
Call customer service center.
Warning then Fault
Main power is not correct
y
Check the main power.
Other problem
Call customer service center.
Warning
Motor temperature is > 100°C
y
See Warning Water flow (
W08
).
Sensor problem
y
Check the sensor connection on
Mot. temp.
connector (see
D 10
).
Other problem
Call customer service center.
Fault
Soft > 1.03
No motor temperature cable
connected
y
In case of ADS 602, check that a
cover plug or a cable is connected
on LI4 logic input (see
D 10
).
y
Check the sensor connection on
Mot. temp.
connector (see
D 10
)
Motor temperature is > 150°C
y
See Warning Water flow (
W08
).
Sensor problem
y
Check the sensor connection on
Mot. temp.
connector (see
D 10
) (and
the Roots temperature sensor on
LI4
on model ADS 602, ADS 1202.)
Other problem
Call customer service center.
Fault
Soft < 1.03
Sensor problem
y
Check the sensor cable connection
on
Mot. temp.
connector (see
D 10
)
Motor temperature is > 150°C
y
See Warning Water flow (
W08
).
Other problem
Call customer service center.
D07 SPEED CONTROL
D07 SPEED CONTROL
W08 WATER FLOW
W08 WATER FLOW
W09/D09 POWER SUPPLY
W09/D09 POWER SUPPLY
W10 MOTOR TEMP.
W10 MOTOR TEMP.
D10 MOTOR TEMP.
D10 MOTOR TEMP.
D10 MOTOR TEMP.
D10 MOTOR TEMP.