2/2
C 70
Use of the inlet isolation valve (accessory)
G 01067 - Edition 02 - Januar
y 10
3. Controlled by the hand held
remode control
4. Controlled via the serial link
For the valve to operate as described below, contact S7 must be
closed (see
B 70
).
Pressing the “
AUX
” button on the hand held remote control opens
or closes the valve, whether the pump is stopped or running (press
the “
ENTER
” button on the hand held remote control to confirm the
operation).
Note: In either local or remote control mode:
– When the pump stops normally or due to an alarm, the valve is
closed and the pump stops as soon as the valve end stop sensor is
triggered or after 3 seconds.
– When the pump is started, the valve opens 3 seconds after the
pump starts.
The valve open control message on the serial link is only taken into
account if contact S7 is closed.
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two