6/7
D 500
The pump is running and then is showing
a warning or fault message
G 03219 - Edition 01 - December 06
Incident
Cause
Troubleshooting
Warning
Pump temperature is < of 15°C to the
temperaturecontrol setting value
Temperature control valve supply
problem
y
Check the temperature control valve
connection on
Therm. ADP
connector
(see
D 10
).
Other problem
Call customer service center.
Warning
LI1 and LI3 logic inputs are not used
to monitor a LL pump
Incorrect monitoring configuration
y
Check that “
LL PUMP OPTION
” is not
valid. (see
C 42
).
Warning
LI1 and LI3 logic inputs are used to
monitor a LL pump
LL pump problem
Call customer service center.
Wiring problem
y
Check the connections on
LI1
and
LI3
connectors (see
D 10
).
Other problem
Call customer service center.
Warning
The maintenance threshold of a
parameter is reached
y
Check maintenance thresholds.
Call customer service center.
Warning
The maintenance threshold of a
parameter is reached
y
Check maintenance thresholds.
Call customer service center.
Warning
The maintenance threshold of a
parameter is reached
y
Check maintenance thresholds.
Call customer service center.
Warning
Warning
The maintenance threshold of a
parameter is reached
y
Check maintenance thresholds.
Call customer service center.
Warning
The maintenance threshold of a
parameter is reached
y
Check maintenance thresholds.
Call customer service center.
Fault on ADS 1202/1802
Incorrect M4 configuration
y
Configure for “ADP 122“ or “ADS“
(see
C 42
).
Other problem
Call customer service center.
Fault
Pump emergency stop is engaged
y
Rotate to unlock the emergency stop
on the front panel display.
Other problem
Call customer service center.
W16 ADP TEMP. LOW
W16 ADP TEMP. LOW
W18 LOAD-LOCK PUMP
W18 LOAD-LOCK PUMP
W18 LOAD-LOCK PUMP
W18 LOAD-LOCK PUMP
W28 LOAD-LOCK MAINT.
W28 LOAD-LOCK MAINT.
W29 ROOTS MAINT.
W29 ROOTS MAINT.
W30 EXHAUST MAINT.
W30 EXHAUST MAINT.
W31 ADP LP MAINT.
W31 ADP LP MAINT.
W32 ADP MAINT.
W32 ADP MAINT.
D34 AR.URGT; CONT.ADP
D34 AR.URGT; CONT.ADP
D35 EMERGENCY STOP
D35 EMERGENCY STOP
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two