G 03219 - Edition 01 - December 06
3/7
D 500
The pump is running and then is showing
a warning or fault message
Incident
Cause
Troubleshooting
Warning or Fault
Analogic input is not used
Incorrect monitoring configuration
y
Check that “
ANALOG INPUT”
menu
is not valid.
Warning or Fault
Analogic input is used
Incorrect monitoring configuration
y
Check the warning or fault
thresholds.
y
Check that the voltage is in
accordance with M4 display.
Other problem
Call customer service center.
Warning or Fault
Logic input is not used
Incorrect monitoring configuration
y
Check that “
LI1 LOG INPUT
” (
LI2
or
LI3
) is not valid.
Warning ou Fault
Logic input is used
Incorrect monitoring configuration
y
Check the warning or fault delays.
y
Check the sensors.
Other problem
Call customer service center.
Warning
Variator option is not used
Incorrect monitoring configuration
y
Check that “
SYSTEM SELECT.
” menu
is properly configurated.
Warning
Roots does not run at nominal speed
Mechanical problem
y
Check the exhaust line.
Other problem
Call customer service center.
Warning
Roots does not run
Motor problem
Call customer service center.
Fault
(ADS 1802 only)
Incorrect monitoring configuration
y
Check that “
SYSTEM SELECT.
” menu
is properly configurated.
W04/D04 ANALOGIC INPUT
W04/D04 ANALOGIC INPUT
W04/D04 ANALOGIC INPUT
W04/D04 ANALOGIC INPUT
W05/D05 LOGIC INPUT 1
W05/D05 LOGIC INPUT 1
W06/D06 LOGIC INPUT 2
W06/D06 LOGIC INPUT 2
W12/D12 LOGIC INPUT 3
W12/D12 LOGIC INPUT 3
W05/D05 LOGIC INPUT 1
W05/D05 LOGIC INPUT 1
W06/D06 LOGIC INPUT 2
W06/D06 LOGIC INPUT 2
W12/D12 LOGIC INPUT 3
W12/D12 LOGIC INPUT 3
W07 SPEED CONTROL
W07 SPEED CONTROL
W07 SPEED CONTROL
W07 SPEED CONTROL
W07 SPEED CONTROL
W07 SPEED CONTROL
D07 SPEED CONTROL
D07 SPEED CONTROL
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two