G 01055 - Edition 07 - Januar
y 10
1/4
B 50
Electrical connection
Customer electrical
installation protection
All the internal electrical connections required for use the pump are
made prior to shipment.
However, the electrical connection of the main power supply is
provided by the user.
Series Two pumps do not include any operator replaceable fuses.
For any change of power supply voltage in relation to the
initialconfiguration, contact the manufacturer’s Support Service.
Main circuit breaker rating (recommended values):
Consump-
tion /
Breaker
rating
200-230V — 50/60 Hz
380-480V — 50/60 Hz
Full-power
consumption
Circuit
breaker
ratting
Full-power
consumption
Circuit
breaker
ratting
ADP 122
9 A
15 A
5 A
10 A
ADS 602
19 A
25 A
11 A
15 A
ADS 1202
25 A
35 A
17 A
25 A
ADS 1802
32 A
40 A
17 A
25 A
The pump is equipped with a 10 000 AIC main breaker.
Electrical shock may result in severe injury.
Always disconnect main power before servicing the pump.
Make sure that main switch is off during electrical connection.
Study the preliminary precautions (see
B 00
).
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two