![Acuster BEAT HF Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/acuster/beat-hf/beat-hf_user-manual_2845321062.webp)
USER MANUAL
BEAT HF
Edition: January 2016
- MAINTENANCE
Revision: No. 4
62
With the cursor on <No> and pressing <START> the display will show the following
screen:
dd/mm/yy
hh:mm
Updating Firmware
Don’t turn off
The backlight of the screen will go off. The update
process lasts about a minute.
5.
When the led of the <START> push-button flashes on and off slowly, click “Start
Process” in the software of the PC.
6.
Once the download is complete, the unit will start up automatically (when the
software is updated with a new version the default language -English- is restored
and the existing fusion records, fusion counter, operator and job data are erased).
7.
Switch off the unit, unplug the transmission cable and verify unit operation.