- MODE OF USE
Revision: No. 4
16
The display backlight will be enabled, an acoustic signal heard and the following
information will appear:
BEAT-HF
GRUPO ACUSTER
S.N. xxxx
V. v.vv
Model:
HF-1=BEAT HF-MV;
HF-2=BEAT HF-2;
BEAT-HF=BEAT HF-Tr.
S.N. xxxx:
unit serial number
V. v.vv:
software version
Once the display time has elapsed (3 seconds), the unit will display the corresponding
screen (depending on the setup and initial messages).
Possible initial messages:
MEMORY FULL (BEAT HF-Tr only)
If the capacity of the memory is already full, the display will show the following message:
dd/mm/yy
hh:mm
Memory full
Fusion impossible
Export or/and eraser fusion records: refer to clause
3.5.3 of this Manual.
TECHNICAL MACHINE SERVICE
If the programmed warning date for the Technical Machine Service in the unit has expired,
the display will show:
dd/mm/yy
hh:mm
Service valid until
dd/mm/yyyy
Confirm
The unit will emit two
acoustic signals and a
message will appear in
the display along with the
date of the next check.
dd/mm/yy
hh:mm
Service valid until
dd/mm/yyyy
If the date dd/mm/yyyy has not expired yet, press <START> to Confirm and proceed
to the next screen (regardless of the existing locking set up of the unit).
If the date dd/mm/yyyy has already expired but the unit is set up without locking, by
pressing <START> you will be able to move to the next screen. However, if the unit is set
up with locking, the display screen won’t show the Confirm option and subsequently the
unit will not be able to be used until the TMS has been carried out.