Warning:
Never change the cartridge or prime the
infusion set
with
an
infusion set inserted
into your body. You risk uncontrolled insulin
delivery. With disconnectable infusion sets, make sure you have discon-
nected the tubing from your site prior to changing or priming.
Always follow the instructions for use of the infusion set you are using.
Warning:
Do not use your insulin pump in hyperbaric chambers and in
hazardous areas of any classification (such as areas where explosive or
flammable gases or vapors could exist) as this might interfere with insulin
delivery and/or lead to harmful situations. Always take off your insulin
pump before entering such areas.
Your insulin pump has not been tested for use in hazardous areas of any classifica-
tion. Always take off your insulin pump before entering such areas.
Warning:
Contact your doctor and/or health care team for an alternate
therapy plan when you interrupt the insulin pump therapy for a longer pe-
riod of time.
Warning:
Before using another insulin pump, always check your person-
al settings before use in order to avoid incorrect insulin delivery. Please
record your personal settings from your current pump and ensure that
they are correctly programmed into the new pump. Make sure that the
time and date are set correctly. If you are not sure about your personal set-
tings or you are less experienced, contact your doctor and/or your health
care team to check your personal settings.
Warning:
Your insulin pump has been developed for use with Accu-Chek
infusion sets. Other infusion sets may not have been tested for compati-
bility with your Accu-Chek D-TRONplus insulin pump and may endanger
your health if used. Infusion sets are
sterile products intended for
single use only.
Sterility is guaranteed for unopened packaging up to
the indicated expiration date.
Do not use if package is damaged.
Accu-Chek D-TRONplus insulin pump Reference Manual
13
Accu-Chek D-TRONplus insulin pump Reference Manual
12
Do not reuse single use materials. Reuse of single use materials might
lead to a malfunction of your insulin pump and to incorrect insulin
delivery and/or an infection. Always handle the items with clean hands.
Warning:
Only put your insulin pump into
RUN
if it is fully and correctly
equipped (including cartridge, adapter and infusion set) and programmed
with your personal settings
approved by your doctor and/or your
health care team.
Warning:
The amount of insulin delivered over a certain period of time
is not limited by your insulin pump. That means several boluses can be set
after the last bolus is delivered.
Warning:
Check both programmed basal rate profiles carefully when you
change from one basal rate profile to the second basal rate profile. If you
are not sure about your personal settings or you are less experienced,
have your doctor and/or your health care team check your personal set-
tings. Work with your doctor and/or your health care team to determine
the timing, amount and type of bolus you need to deliver. Make sure you
know your personal insulin/carbohydrate ratio and your correction bolus
ratio.
Warning:
Training to use and using your Accu-Chek D-TRONplus insulin
pump requires the support of an experienced doctor and/or your health
care team. Regular visits with your doctor and/or your health care team
are absolutely essential during insulin pump therapy. Only change your
personal settings after consulting your doctor and/or your health care
team. Always follow the instructions given by your doctor and/or your
health care team.
Successful pump therapy requires frequent self-monitoring of blood glucose levels, at
least four times a day, or as directed by your health care team.