Accu-Chek D-TRONplus insulin pump Reference Manual
51
Accu-Chek D-TRONplus insulin pump Reference Manual
50
Caution:
If you discover incomplete characters, numbers or symbols on
the display, your insulin pump needs to be returned since you may not be
notified of any changes of your insulin pump in a timely manner. Please
contact your Accu-Chek customer service center for insulin pumps.
Caution:
Do not interrupt the start-up procedure by pressing keys or in-
serting a cartridge or adapter. Interruption of the start-up procedure may
lead to malfunctions of your insulin pump.
Caution:
If your insulin pump does not beep and vibrate as described, or
if individual segments are shown incompletely or not at all, your insulin
pump needs to be returned since you may not be notified of any changes
of your insulin pump in a timely manner. Please contact your Accu-Chek
customer service center for insulin pumps.
Info:
If the PowerPack has been removed from your insulin pump for more
than 1 hour, an “Alert A3: Set Time/Date” might occur after the PowerPack
has been inserted. Reset the time and date. Your insulin pump configura-
tion and settings (basal rate, active profile, automatic off, etc.) and the
event memory (bolus history, history of daily totals, history of alerts and
errors) remain, irrespective of the PowerPack voltage and the time your
insulin pump has been without a PowerPack.
3.1.2 Change the PowerPack
1. Make sure your Accu-Chek D-TRONplus insulin pump
is in
STOP
.
2. Remove or disconnect your infusion set from the site.
After inserting the PowerPack, your insulin pump performs the following start-up pro-
cedure:
• The display indicates the software version (i.e. 3.00) and 77.
Your insulin pump beeps a sequence of three and two short
beeps.
• The display shows the internal PowerPack test (1, 2 or 3).
There is no specific meaning related to these figures or any
action needed dependent on the figure displayed.
• If the remaining cartridge content is lost from the memory,
the piston rod returns.
Your insulin pump then performs the following self-test:
• All segments are shown on the display.
Your insulin pump beeps a sequence of three and two short
beeps and vibrates once.
Caution:
If during the “pump self test” your in-
sulin pump does not beep a sequence of three
and two short beeps and/or does not vibrate
once, please contact your Accu-Chek customer
service center for insulin pumps.
• The numbers in the three number groups go from 0000 to
9999.
Check if the numbers appear completely.Your insulin pump
beeps a melody to inform you that the start-up procedure
has ended.
• Your insulin pump returns to
STOP
. If your insulin pump is
left in
STOP
, the
STOP
-Warning is given every minute.