Accu-Chek D-TRONplus insulin pump Reference Manual
107
Accu-Chek D-TRONplus insulin pump Reference Manual
106
5. Remove the infusion set from the adapter and dispose of it
properly.
You may use the battery tool to unscrew the luer connection
between the infusion set and adapter if it cannot be un-
screwed by hand.
Warning:
Your insulin pump has been designed for use with Accu-Chek
infusion sets only. The use of other infusion sets can endanger your health
and may void the warranty.
Infusion sets are sterile products
intended for single use only.
Sterility is guaranteed for unopened
packaging up to the indicated expiration date.
Do not use sterile
products if the package is damaged.
Do not reuse single use materials. Reuse of single use materials might
lead to a malfunction of your insulin pump and to incorrect insulin de-
livery and/or an infection. Always handle the items with clean hands.
6. Set up your insulin pump with the new infusion set.
Please refer to section 4 “Prepare the cartridge and attach the
adapter and infusion set” for further information.
Caution:
If the luer connection between the infusion set and adapter can
not be unscrewed by hand, it is recommended that you use the battery
tool. The battery tool has a notch that fits most infusion set luers. Use the
battery tool for unscrewing only. Never use it for attaching the infusion set.
To change the infusion set only
1. Have the following materials ready:
• Your insulin pump
• A new infusion set
• New site dressing, if desired
• An antiseptic wipe
2. Make sure your insulin pump is in
STOP
.
3. Remove your used infusion set from the site.
4. Rotate the adapter into the replace position.