Using the Z Vent Ventilator
4-34
www.zoll.com
9650-002360-01 Rev. A
Heart Rate Limit Conflict
Popup message triggers when the user attempts to set the HR Low limit > the HR High limit.
Message: Cannot adjust low limit higher than high limit
PS Conflict
Popup message triggers when the user attempts to set the PS > the PIP High limit - PEEP
pressure.
Message: Cannot adjust PS higher than PIP High Limit - PEEP
Leak Comp.
Popup message triggers when the user attempts to initiate leak compensation (LC). To do this, the
user must press the
Accept
button and then select LC followed by
Accept
again to confirm the
setting change.
Message: Some Alarms Disabled! Configure Alarms for Patient!
Mode Conflict
Popup message triggers when the user attempts to initiate leak compensation (LC) during volume
targeted ventilation. Note: LC is only available during pressure targeted ventilation.
Message: Cannot select Volume targeted control breaths with Leak Compensation on -- first turn
Leak Compensation off
Inverse I:E
Popup message triggers when the user attempts to set an inverse I:E ratio. To do this, the user
must press the
Accept
button and then adjust the I:E ratio to the desired inverse value and press
Accept
again to confirm the setting change.
Message: Confirmation required -- press accept key to allow inverse I:E
Excessive Volume for Infant Circuit
Popup message triggers when the user attempts to set a Vt > 300 ml when the tubing compliance
compensation is set to Pediatric.
Message: Press accept to confirm use of adult circuit
Insufficient Volume for Adult/Ped Circuit
Popup message triggers when the user attempts to set a Vt < 200 ml when the tubing compliance
compensation is set to Adult.
Message: Press accept to confirm use of infant circuit
High PEEP Setting
Popup message triggers during CPAP mode when the user attempts to set the PEEP > 15 cm H
2
O.
To do this, the user must press the
Accept
button and then adjust PEEP to the desired value and
press
Accept
again to confirm the setting change.
Message: Confirmation required -- press accept key to allow PEEP above 15
High EPAP Setting
Popup message triggers during BL mode when the user attempts to set the EPAP > 15 cm H
2
O. To
do this, the user must press the
Accept
button and then adjust PEEP to the desired value and
press
Accept
again to confirm the setting change.
Message: Confirmation required -- press accept key to allow EPAP above 15