Product Overview
2-16
www.zoll.com
9650-002360-01 Rev. A
Infant/Pediatric, 12 ft Patient Circuit
The Infant/Pediatric, 12 ft patient circuit has the following specifications:
•
Internal Diameter: 10 mm
•
Inspiratory Resistance: R
INSP
@ 15 Lpm: 0.17 hPa/l/min*
•
Expiratory Resistance: R
EXP
@ 15 Lpm: 0.17 hPa/l/min
•
Tubing Compliance: C
T
@ 60 hPa: 0.8 ml/hPa
•
Dead Space: 4.2 ml
•
Maximum Working Pressure: 100 hPa (cm H
2
O)
Note:
The extended length of the tubing in the 12 ft circuit results in a higher RINSP
compared to the 6 ft circuit.
Warning!
Compressible volume can significantly decrease the delivered tidal volume. When
managing patients at risk, always correct for compressible volume Use the Vt Context
Menu to adjust Tubing Compliance and Compressible Volume Measurements.
Warning!
Do not use the 12 ft circuit with PEEP settings below 5 cm H
2
O.
Warning!
Given the additional length of the 12ft circuit, the system may not be able to trap PEEP
in patients with short expiatory time. Always ensure that the device is performing as
required.