Using the Z Vent Ventilator
4-40
www.zoll.com
9650-002360-01 Rev. A
To view unit information for the ventilator, follow these steps:
1. Press the
Menu
button to display the Menu.
2.
Turn the
Dial
to highlight
Unit Info
, and then press the
Accept
button. The list of information
displays with
Use Statistics
highlighted.
3.
Press the
Accept
button
to display ventilator usage in hours and minutes (
hh:mm
)
4.
Press the
Mute/Cancel
button to return to the Menu.
Alarm History
The Alarm History provides a list of the alarm messages generated by the ventilator during use.
The history holds a maximum of 256 events.
Note:
Power cycling does not clear the alarm history.
Each list entry provides the following information
•
The Alarm Name
•
The Service Code
•
The calendar date when the event was entered
•
Whether the alarm is set or cleared
•
The time (24hr) when the event was entered
•
The log entry number <
NNN
> The higher the number the more recent the alarm message
event was logged
To display the Alarm History, follow these steps:
1. Press the
Menu
button to display the Menu.
2.
Turn the
Dial
to highlight
Alarm History
, and then press the
Accept
button. The most recent
history entry displays.
3.
Turn the
Dial
counterclockwise to view earlier entries. Turn the
Dial
clockwise to view more
recent entries.
4.
Press the
Mute/Cancel
button to return to the Menu.
5.
Press the
Mute/Cancel
button again to return to the user interface.
The alarm condition is not logged when alarm is active but not displayed (due to other higher
priority alarm indication as explained in the Alarms Chapter of this manual), or if the alarm is
disabled in the alarm configuration the condition is not logged.