Setting Up the Z Vent Ventilator
9650-002360-01 Rev. A
ZOLL Z Vent Ventilator Operator’s Guide
3-5
Never use O
2
flows > 10-12 liters/min. Flows greater than this can cause the baseline
pressure to drift, waste oxygen, and may cause an INCOMPLETE EXHALATION
alarm.
Warning!
Always monitor the patient's oxygenation using a pulse oximeter. The O
2
flow from a
concentrator or other O
2
source may not be adequate to achieve the desired SPO
2
target. Failure to follow the Instructions and WARNINGS provided with the O
2
Reservoir could result in an adverse effect on the patient.
Note:
Due to the slight difference between the densities of air and O
2
, the tidal volume will
decrease slightly as O
2
is entrained. The worst case is a < 10% decrease in tidal
volume when the entrained O
2
results in FIO
2
of 100%.
The table below shows both the affect on tidal volume and the resultant FIO
2
supply
rate.
Bacterial/Viral (BV) Filter
If the ventilator will operate in an environment where the patient is at risk from cross
contamination or airborne pathogens, you may choose to attach a BV filter (See Chapter 6,
“Operating Environments” for more information on this filter).
Chemical/Biological C2A1 Filter
If the ventilator will operate in a contaminated environment, you may choose to attach a
chemical/biological C2A1 filter obtained from a Chemical/Biological Filter supplier.
Z Vent
AC 12, Vt 700, PEEP 5, I:E 1:2.5
O
2
Flow
0
1
2
3
4
5
6
7
8
FIO
2
21
30
38
48
57
70
80
89
100
Vt (set)
740
732
725
718
711
703
691
689
682
Vt (actual)
700
692
685
678
671
663
651
649
642
% Chg
0
-1.1
-2.1
-3.1
-4.1
-5.3
-7.0
-7.3
-8.3
AC 12, Vt 500, PEEP 5, I:E 1:2.5
O
2
Flow
0
1
2
3
4
5
6
FIO
2
21
30
43
56
69
89
100
Vt (set)
527
523
514
506
502
493
486
Vt (actual)
500
496
487
479
475
466
459
% Chg
0
-0.8
-2.6
-4.2
-5.0
-6.8
-8.2
AC 18, Vt300, PEEP 5, I:E 1:2.5
O
2
Flow
0
1
2
3
4
5
6
FIO
2
21
32
47
62
76
96
100
Vt (set)
312
307
303
299
298
291
287
Vt (actual)
300
295
291
287
286
279
275
% Chg
0
-1.7
-3.0
-4.3
-4.7
-7.0
-8.3