9650-002360-01 Rev. A
ZOLL Z Vent Ventilator Operator’s Guide
A-1
Appendix A
Specifications
General
Parameter
Operating Range
Operating Modes
AC,SIMV, CPAP with and without Pressure Support, BL Mode, and Leak
Compensation for Invasive and Noninvasive ventilation
Breath Target
Volume or Pressure
Flow Rate
0 to 100 LPM at 40 cm H
2
O
Breath Rate
1 to 80 BPM +/- 1 BPM over the interval
Setting Resolution: 1 BPM
Measurement: 1 to 90 BPM +/- 1 BPM over the interval
Inspiratory Time (Ti)
Setting:
0 to 3 +/- 0.1 sec for I:E from 1:1 to 1:99
0 to 5 +/- 0.1 sec for I:E from 4.0:1 to 1:99 (Inverse I:E)
Setting resolution: 0.05 s
Tidal Volume
Setting: 50 to 2000 ml ATPD +/- (5 ml +10% setting) Inverse I:E ratio is available.
Setting Resolution: 10 ml (Measurement: 0 to 9999 ml ATPD +/- (5 ml +10% setting)
above 100 ml (5 ml below 100 ml)
FIO
2
21 to 100% +/- (3% of full scale +/- 10% of setting)
PEEP/EPAP
Setting: 0 to 30 cm H
2
O+/- (2 cm H
2
O + 8% of reading)
Setting Resolution: 1 cm H
2
O
Measurement: 0 to 30 cm H
2
O +/- (2 cm H
2
O + 8% of reading)
Peak Inspiratory Pressure
(PIP)
Setting: 10 to 80 cm H
2
O +/- (2 cm H
2
O + 8% of setting)
Setting Resolution: 1 cm H
2
O
Measurement: 0 to 99 cm H
2
O +/- (2 cm H
2
O + 8% of reading)
Pressure Support (PS)/
IPAP
0 to 60 cm H
2
O +/- (2 cm H
2
O + 8% of setting).