Using the Z Vent Ventilator
4-6
www.zoll.com
9650-002360-01 Rev. A
Changing a Parameter Setting
To change a parameter, follow these steps.
1. Access the parameter you intend to change using the parameter buttons as described in the
previous sections.
2.
Press the
Accept
button to highlight your choice. Multiple presses are required for secondary
parameters.
3.
Turn the
Dial
to adjust the parameter value
4.
Press the
Accept
button to enable your choice.
The examples that follow describe how to change a primary parameter, secondary parameter,
and a context menu parameter.
Example 1 — Changing a Primary Parameter
In Example 1, the ventilator is operating with the default Adult parameters. The user wants to
change the Mode from AC (Adult Default) to CPAP.
To change the Mode parameter from AC to CPAP, follow these steps:
1. Press the
Mode
parameter button once.
2. Turn the
Dial
until the Mode parameter window displays CPAP.
3. Press the
Accept
button.
Example 2 — Changing a Secondary Parameter
In Example 2, the ventilator is operating with the default Adult parameters and user wants to
adjust the PEEP secondary parameter setting from 5 (the adult default setting) to 7 cm H
2
O.
To change the secondary parameter PEEP setting from 5 to 7, follow these steps:
1. Press the
PIP
parameter button until PEEP is highlighted.
2. Turn the
Dial
until the PEEP setting is 7 as displayed in the PIP parameter window.
3. Press the
Accept
button.
Example 3— Changing a Context Menu Parameter
In Example 3, the ventilator is operating with the default Adult parameters and user wants to
change the Masimo Pulse Oximeter Sensitivity from Norm (normal) which is the adult default
setting – to Max (maximum).
To change the Sensitivity setting from
Norm
to
Max
, follow these steps:
1. Press and hold the
HR
or
SpO
2
parameter button until the Masimo context menu displays.
2.
Release the parameter button, then turn the
Dial
until the Sensitivity choice highlights, then
press the
Accept
button. The Sensitivity setting highlights.
3.
Turn the
Dial
until the setting changes to Max (maximum).
4.
Press the
Accept
button.