Alarms
5-34
www.zoll.com
9650-002360-01 Rev. A
3092
Inspiratory Demand
Alarm triggers when the end-inspiratory pressure is < -1.0 cm H
2
O for 3
consecutive breaths. This can occur due to changes in the patient’s status,
where the patient attempts to inhale more gas than what is currently set. When
this occurs, the user should note if the patient is breathing or fighting with the
ventilator. The user should increase the flow rate (by decreasing the inspiratory
time) and/or reduce the rise time. The physician should be consulted.
Mitigation/Info: Patient May be Breathing with the Ventilator, Increase I
Time and/or Decrease Rise Time, Check Patient and Circuit for Leaks,
Disable Alarm, **Consult Physician**
3110
RTC Battery Low
Alarm triggers when the real-time clock (RTC) battery is < ~2.5 VDC. The alarm
condition is checked at start up and if this alarm occurs the device is safe to
operate but the user should replace the device when appropriate and consult the
service center for additional information. The user cannot change the RTC
battery. The RTC battery provides power for the clock that tracks the local time. It
is replaced every 4 years during preventive maintenance.
Mitigation/Info: Vent Fully Functional, **Contact Service Center**
3120
PM Due
Alarm triggers at start up when the preselected number of days has elapsed
since the last calibration. When appropriate the device should be replaced and
sent for preventive maintenance. The low priority message serves as a reminder.
Calibration is due every 365 days or 730 days for devices configured for
stockpile use (check with your organization regarding the configuration of your
device). Users should schedule the device for service as soon as possible. Users
can suspend the yellow alarm notification for the current use by turning the alarm
off using the Alarm Configuration submenu in the Menu.
Mitigation/Info: Preventive Maintenance Due, Ventilator Functioning with
No Faults, **When Appropriate, Contact Service Center **
3121
Power Cycle Need
This alarm occurs when the device has been running continuously for 30 days. In
order to check the flow pneumotach, which is done a startup, the user should
manually ventilate the patient and cycle the power. Once this is done the user
can select the Last Settings options from the Start Menu and continue operation
if not faults are detected during the self check. If nonoperating alarms occur,
contact the service center for additional information.
Mitigation/Info: Power-on Self Checks are Due, When Appropriate,
Power Off Then On, Verify Proper Settings, **Review Manual For
Additional Information**
Service Code
Alarm Name/Mitigation/Resolution