Alarms
9650-002360-01 Rev. A
ZOLL Z Vent Ventilator Operator’s Guide
5-5
Alarm Title (units)
Adjustable Range, Default
Service Code
Airway Pressure High
PIP High (cm H
2
0)
20 to 100 (>60 requires confirmation)
Default: Adult:35 Pediatric: 30
2070
Low Airway Pressure
PIP Low (cm H
2
0)
3 to 35, OFF (with popup confirmation; 3
without confirmation)
Default: 10 in AC, SIMV
Default: OFF in CPAP or BL
See transition and feature notes in
Chapter 4.
2071
High Tidal Volume
Vt High
(ml)
0 to 2000, OFF (with popup confirmation;
2000 without confirmation)
Default: Adult: N/A Pediatric: 500
Default: OFF in CPAP or BL
See transition and feature notes in
Chapter 4.
2072
Low Tidal Volume
Vt Low
(ml)
0 to 2000, OFF (with popup confirmation;
0 without confirmation)
Default: Adult: N/A Pediatric: 500
Default: OFF in CPAP or BL
See transition and feature notes in
Chapter 4.
2073
High Breath Rate
(breath per minute)
20 to 99, OFF (with popup confirmation;
99 without confirmation)
Default: Adult: 25 Pediatric: 30
2074
Low Breath Rate/Apnea
2 to 25 (with popup confirmation below 6)
Default: Adult: 8 Pediatric: 10
2075
Apnea
2076
PEEP Leak
Not adjustable, OFF (with Alarm Menu)
2090
Insufficient Flow
Not adjustable
2095
Spontaneous Breath-PIP High
20 to 100
Default = Adult:35 Pediatric: 30
2170
Spontaneous Breath-PIP Low
3 to 35, OFF (with popup confirmation; 3
without confirmation)
Default: 10 in AC, SIMV
Default: OFF in CPAP or BL
See transition and feature notes in
Chapter 4.
2171