Using the Z Vent Ventilator
4-2
www.zoll.com
9650-002360-01 Rev. A
Initial Operation with Default Parameter Settings
Following power on, the ventilator goes through a Self-Check (a set of system tests and
checks). If the Self-Check passes, the LED array turns green and the Start Menu displays,
indicating that the ventilator is operational.
The Start Menu enables the user to choose between predefined ventilator parameter settings
(for adult patients, pediatric patients, patients requiring Mask CPAP), a previously saved set of
Custom parameter settings, or the parameter settings last used during ventilator operation. The
Start Menu choices include:
Warning!
Do not connect patient to the ventilator while the Start Menu is active.
Default Parameter Settings for Adult, Pediatric and Mask CPAP
The default settings for Adult, Pediatric, and Mask CPAP are as follows:
Adult Default Parameter Setting Values
Choice
Description
Adult
Preset ventilation parameter settings for adult patients.
Pediatric
Preset ventilation parameter settings for pediatric patients.
Mask CPAP
Preset ventilation parameter settings for Mask CPAP (Continuous
Positive Airway Pressure) ventilation.
Custom
Ventilation parameter settings previously saved by a user.
Last Settings
The settings enabled on the ventilator during its last use (but not
saved as Custom Settings by the user).
Parameter
Default Setting Value
Mode
AC (V)
BPM
12
I:E
1:3
VT
450
PEEP
5
PIP Limit
35
FIO
2
21