Using the Z Vent Ventilator
9650-002360-01 Rev. A
ZOLL Z Vent Ventilator Operator’s Guide
4-5
Figure 4-2 Parameter Buttons
Navigating the Parameter Windows Using Parameter Buttons
Parameter values are accessed with a parameter button as follows:
•
Single Press:
highlights the primary parameter for the chosen parameter window
•
Multiple Presses:
highlights secondary parameters and alarm limits. (Multiple presses
highlight secondary parameters moving in a clockwise direction)
•
Press and Hold:
opens the context menu (for those primary parameters that have a context
menu). Turning the
Dial
highlights context menu items.
Note:
If you attempt to set parameter values that are outside the typical clinical range of
settings, the ventilator displays popup messages that ask if you are sure you would like
to set the parameter to that value. We describe popup messages in more detail later in
this chapter. To set the parameter above the limit, you must press the
Accept
button,
and then adjust the values and press the
Accept
button again.