Product Overview
9650-002360-01 Rev. A
ZOLL Z Vent Ventilator Operator’s Guide
2-13
Intended Use
The pediatric/adult patient circuits are intended for use when delivering tidal volume from
200 ml to Adult.
The infant/pediatric patient circuit is intended for use when delivering tidal volume from
50 ml to 300 ml.
Warning!
Patient circuits are non-sterile and intended for Single Patient Use Only
Caution
During use the circuit may come into contact with biohazard material. Handle carefully to avoid
cross-contamination.
Not intended for use with heated humidifier.
Note:
ZOLL Medical Corporation recommends that you examine the patient circuit on a
daily basis for damage or wear, such as cracking, discoloration, or disfigurement. If
there is any sign of physical degradation or if the ventilator has patient circuit alarm
conditions replace the patient circuit.
Use of Heat and Moisture Exchangers
Heat and Moisture Exchangers (HMEs) can be used with the device. The HME provides heat
and moisture to the inspired gas by recycling the heat and moisture contained in the patient's
exhaled gas. While HMEs may not be suitable for all applications, they facilitate portability in
a way that conventional humidifiers cannot. The device can be used with an optional HME or
an optional HME/bacterial viral filter (HMEF). Be sure to follow all instructions provided by
the manufacturer.
Note:
Use of the HME will cause a slight increase in the inspiratory and expiratory
resistance. Always monitor the patient and adjust the ventilator as needed.
ZOLL does not offer a heated humidification option for the device.
Warning!
Users should use the appropriate HME for the patient's size. Failure to do so can result
in excessive dead space and lead to hypercapnia and hypoxia.