Alarms
5-20
www.zoll.com
9650-002360-01 Rev. A
1175
Self Check Failure - Internal COMM
Alarm triggers when the internal communication bus and the host are not able to
communicate with the subassemblies. If this failure occurs, the user should
manually ventilate the patient, replace the ventilator and contact the service
center for additional information.
Mitigation/Infor: Internal COMM, Manually Ventilate Patient, **Contact
Service Center**
1176
Self Check Failure - Internal COMM
Alarm triggers when the calibration file fails its integrity check. The user should
manually ventilate the patient, replace the ventilator and contact the service
center for additional information.
Mitigation/Info: Internal COMM, Manually Ventilate Patient, **Contact
Service Center**
1420
Self Check: Complete Power Failure
Alarm triggers when power is lost from both the internal battery and an external
source during operation. When this occurs, the LCD blanks (no power for
operation), the audible alarm pulses rapidly, and the visual alarm flashes rapidly.
This alarm will last approximately two minutes. If the device can be recharged
after the failure and there are no other issues it can be returned to service. If
there are any questions, contact the service center for additional information.
Mitigation/Info: No LCD Display
1430
Drained Battery
Alarm triggers when the internal battery power drops below the amount required
to provide ventilation and external power is not connected. When this occurs
there is enough power to operate the user interface and provide information to
the user. The user should be manually ventilate the patient while an external
source of power is sought. To cancel the alarm and begin operation with external
power the device must be turned off and then back on.
Mitigation/Info: Manually Ventilate Patient, Connect External Power,
**Contact Service Center**
1471
Self Check Failure - Internal COMM
Alarm triggers when the device is no longer able to communicate with the User
Interface Module (UIM) and the interface controls. When this occurs ventilation
continues at the current settings or the backup mode settings and the high
priority alarm sounds. The user should manually ventilate the patient, replace the
ventilator and contact the service center for additional information.
Mitigation/Info: Internal COMM, Manually Ventilate Patient, **Contact
Service Center**
Service Code
Alarm Name/Mitigation/Resolution