Version 6.2
G-30-1767-en
Page 67
INTRABEAM Water Phantom
Care and maintenance
Care and maintenance
Safety inspection
To prevent any impairment of the system's safety as a result of aging, wear
etc., the operating company must ensure, in accordance with the applicable
national regulations, that the regular technical safety checks defined for this
system are performed on schedule and to the stipulated extent.
The safety checks must be performed only by the manufacturer or qualified
personnel.
The scope of the safety checks of the system should include at least the fol-
lowing points:
–
Availability of the Instructions for Use
–
Visual inspection of the product for damage, as well as legibility of the la-
bels
–
Functional check of all adjusting knobs.
Calibration of the IC ionization chamber
Note
Incorrect measuring results may be obtained if you do not have the IC ioni-
zation chamber calibrated at the stipulated intervals.
•
Observe the calibration intervals specified by the manufacturer of the IC
ionization chamber and comply with the relevant requirements applica-
ble in your country.
Summary of Contents for INTRABEAM
Page 1: ...ZEISS INTRABEAM Water Phantom Instructions for Use G 30 1767 en Version 6 2 2022 02 24...
Page 4: ...INTRABEAM Water Phantom Version 6 2 Page 4 G 30 1767 en Blank page for your notes...
Page 45: ...Version 6 2 G 30 1767 en Page 45 INTRABEAM Water Phantom Operation Blank page for your notes...
Page 47: ...Version 6 2 G 30 1767 en Page 47 INTRABEAM Water Phantom Operation...
Page 49: ...Version 6 2 G 30 1767 en Page 49 INTRABEAM Water Phantom Operation Blank page for your notes...
Page 74: ...System data INTRABEAM Water Phantom Version 6 2 Page 74 G 30 1767 en Blank page for your notes...
Page 80: ...Index INTRABEAM Water Phantom Version 6 2 Page 80 G 30 1767 en Blank page for your notes...
Page 81: ...Version 6 2 G 30 1767 en Page 81 INTRABEAM Water Phantom Blank page for your notes...