Operation
INTRABEAM Water Phantom
Version 6.2
Page 46
G-30-1767-en
INTRABEAM 600
The measurement procedure for determining the depth dose
for the INTRABEAM 600 is explained below.
Measuring the depth dose with the INTRABEAM 600
Starting the measurement of the depth dose curve (INTRABEAM 600)
You will find detailed explanations of how to operate the software in the In-
structions for Use of your INTRABEAM 600.
You may already have performed the following adjustments in order to ex-
actly position the probe tip vertically by checking its position via current meas-
urement.
•
Open the "Treatment prescription" window.
•
Create a "patient", for example "XRS_test_1".
•
Make the settings for the XRS 4:
40
μ
A beam current. 50 kV acceleration voltage, no applicator
•
Select the following settings in order to obtain a treatment duration of
several hours:
Prescribed dose 90 Gy, treatment depth 30 mm
•
Start the "treatment" and wait until the display for the internal X-ray mon-
itor indicates approx. 100%.
Summary of Contents for INTRABEAM
Page 1: ...ZEISS INTRABEAM Water Phantom Instructions for Use G 30 1767 en Version 6 2 2022 02 24...
Page 4: ...INTRABEAM Water Phantom Version 6 2 Page 4 G 30 1767 en Blank page for your notes...
Page 45: ...Version 6 2 G 30 1767 en Page 45 INTRABEAM Water Phantom Operation Blank page for your notes...
Page 47: ...Version 6 2 G 30 1767 en Page 47 INTRABEAM Water Phantom Operation...
Page 49: ...Version 6 2 G 30 1767 en Page 49 INTRABEAM Water Phantom Operation Blank page for your notes...
Page 74: ...System data INTRABEAM Water Phantom Version 6 2 Page 74 G 30 1767 en Blank page for your notes...
Page 80: ...Index INTRABEAM Water Phantom Version 6 2 Page 80 G 30 1767 en Blank page for your notes...
Page 81: ...Version 6 2 G 30 1767 en Page 81 INTRABEAM Water Phantom Blank page for your notes...