Operation
INTRABEAM Water Phantom
Version 6.2
Page 58
G-30-1767-en
Measuring isotropy
If air bubbles have accumulated on the probe tip, you must remove them prior
to the measurement, as they may negatively influence the measurement re-
sult:
•
Use a longer spatula or a plastic or glass rod to wipe off the air bubbles.
Calculating deviations
The respective deviation of the 8 measured values from their mean value di-
rectly results in the deviation from the ideal spherical symmetry in the X/Y
plane.
Measuring isotropy with spherical applicator
The geometry of the spherical radiation emission of the probe tip is checked
on an XRS 4 when measuring the isotropy. The isotropy check is performed at
maximum beam current and the beam voltage selected for the respective XRS
4.
Positioning the applicator with the XRS probe for isotropy measurement
As a rule, the isotropy of the X-ray probe is initially recorded without an ap-
plicator. It therefore can be assumed that there will be no need to completely
reposition the probe tip when measuring the isotropy with a spherical appli-
cator.
Exact positioning of the spherical applicator with the X-ray probe is a precon-
dition for accurate and meaningful measuring results.
NOTE
If the user does not rotate the XRS 4 through a full 360° revolution, measured
values for the isotropy will be missing.
•
Measure in all eight detent positions in order enable determination of the
isotropy in the entire X,Y plane.
Summary of Contents for INTRABEAM
Page 1: ...ZEISS INTRABEAM Water Phantom Instructions for Use G 30 1767 en Version 6 2 2022 02 24...
Page 4: ...INTRABEAM Water Phantom Version 6 2 Page 4 G 30 1767 en Blank page for your notes...
Page 45: ...Version 6 2 G 30 1767 en Page 45 INTRABEAM Water Phantom Operation Blank page for your notes...
Page 47: ...Version 6 2 G 30 1767 en Page 47 INTRABEAM Water Phantom Operation...
Page 49: ...Version 6 2 G 30 1767 en Page 49 INTRABEAM Water Phantom Operation Blank page for your notes...
Page 74: ...System data INTRABEAM Water Phantom Version 6 2 Page 74 G 30 1767 en Blank page for your notes...
Page 80: ...Index INTRABEAM Water Phantom Version 6 2 Page 80 G 30 1767 en Blank page for your notes...
Page 81: ...Version 6 2 G 30 1767 en Page 81 INTRABEAM Water Phantom Blank page for your notes...