Version 6.2
G-30-1767-en
Page 53
INTRABEAM Water Phantom
Operation
Measuring the depth dose curve with spherical applicator
Using the depth dose curve, a characteristic curve from the dose rate [Gy/min]
in relation to the penetration depth [mm] of the respective XRS 4 is measured
at 40 kV and at 50 kV.
As a rule, the depth dose of the X-ray probe is initially recorded without an
applicator. It therefore can be assumed that there will be no need to com-
pletely reposition the probe tip when measuring the depth dose curve with a
spherical applicator.
Inserting the spherical applicator in the INTRABEAM Water Phantom
Exact positioning of the applicator (with the probe tip) is a precondition for
accurate and meaningful measuring results.
•
Do not adjust the positioning unit in the X and Y axes.
•
Unless this has already been done:Note down the previously determined
value of Z
1
[mm], see page 41.
•
Release the locking lever on the adjusting knob for the Z axis.
•
Place the positioning unit all the way up in the Z axis.
•
Pull the two covers of the water tank apart sideways over approx. 1/3 of
their surface.
•
Loosen the screws for fixing the XRS mount in the rotary feedthrough.
•
Remove the XRS mount.
NOTE
Contamination of the contact surfaces between the adapter plate and the ro-
tary feedthrough may result in uneven seating of the XRS 4, thus causing sig-
nificant deviations of the measured depth dose curve from the calibration
file.
•
Make sure that these contact surfaces are always clean.
Summary of Contents for INTRABEAM
Page 1: ...ZEISS INTRABEAM Water Phantom Instructions for Use G 30 1767 en Version 6 2 2022 02 24...
Page 4: ...INTRABEAM Water Phantom Version 6 2 Page 4 G 30 1767 en Blank page for your notes...
Page 45: ...Version 6 2 G 30 1767 en Page 45 INTRABEAM Water Phantom Operation Blank page for your notes...
Page 47: ...Version 6 2 G 30 1767 en Page 47 INTRABEAM Water Phantom Operation...
Page 49: ...Version 6 2 G 30 1767 en Page 49 INTRABEAM Water Phantom Operation Blank page for your notes...
Page 74: ...System data INTRABEAM Water Phantom Version 6 2 Page 74 G 30 1767 en Blank page for your notes...
Page 80: ...Index INTRABEAM Water Phantom Version 6 2 Page 80 G 30 1767 en Blank page for your notes...
Page 81: ...Version 6 2 G 30 1767 en Page 81 INTRABEAM Water Phantom Blank page for your notes...