![background image](http://html2.mh-extra.com/html/zeiss/axioscope-5/axioscope-5_instruction-manual_3365198195.webp)
ZEISS
10 Accessories and System Expansions | 10.16 Assembling the Intermediate Plate for Analyzer Slider
10.16 Assembling the Intermediate Plate for Analyzer Slider
1
2
3
4
5
6
7
Fig. 107: Installing the intermediate plate for analyzer slider
1
Clamping screw of the intermediate
plate
2
Stand mount
3
Clamping screw of the stand
4
Binocular tube
5
Intermediate plate
6
Mount of the intermediate plate
7
Slider slot
Parts and Tools
Hex Key, 3.0 mm
Procedure
1. Loosen the clamping screw of the stand
3
.
2. Replace the binocular tube
4
.
3. Unscrew the tube lens of the tube (accessible from the bottom). Use the included ring tool.
4. Screw the tube lens included with the intermediate plate into the binocular tube.
5. Insert the intermediate plate
5
with the dovetail ring into the stand mount
2
.
6. Tighten the clamping screw of the stand .
7. Hold the binocular tube at an angle, insert it with the dovetail ring into the mount of the in-
termediate plate
6
and turn into a horizontal position.
8. Rotate the binocular tube into the desired observation position.
9. Tighten the clamping screw of the intermediate plate
1
.
10. Insert the analyzer slider into the slider slot
7
, if applicable.
Proceed in the reverse order for removal.
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001