6 Care and Maintenance | 6.3 Maintenance Work
6.3.9 Exchanging the Fuses in the External Power Supply for HAL 100
DANGER
Electric shock due to live parts
When the external power supply is still switched on, coming in contact with live parts can lead
to electric shock or burn.
4
Switch off the external power supply prior to opening or cleaning.
4
Disconnect the external power supply from the mains.
2
1
3
1
Fuse
2
Fuse holder
3
Fuse compartment
Parts and Tools
2x Fuse type T5.0 A/H/250V 5x20 mm
Prerequisite
ü
The microscope is switched off and disconnected from the mains.
Procedure
1. If the fuses fail, first check the cause and remedy technical problems properly.
2. Remove the fuse holder
2
on the rear side of the stand.
3. Remove fuses
1
from the fuse holder.
4. Insert new fuses.
5. Push the fuse holder back into the fuse compartment
3
until it locks in place.
6. Bring the microscope back into operation.
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001